Accuracy and Precision of Uric Acid Examination Point of Care Testing Method and Uricase Enzymatic Colorimetric Method

Arini, Fitria Yulfrida and Handayati, Anik and Puspitasari, Ayu (2021) Accuracy and Precision of Uric Acid Examination Point of Care Testing Method and Uricase Enzymatic Colorimetric Method. Prooceding ICoHPS, 1 (1). pp. 34-38.

[thumbnail of Proceeding_ICOMLT_Ayu_2021.pdf.pdf] Text
Proceeding_ICOMLT_Ayu_2021.pdf.pdf

Download (1MB)

Abstract

The method that often used in clinical chemistry examinations in laboratory such as analysis of blood glucose, uric acid, cholesterol, triglycerides, bilirubin, and albumin are POCT method and the uricase enzymatic colorimetric method. The POCT method is a automatic method that quickly obtains result. While the uricase enzymatic colorimetric method is a manual method that still requires special treatments to obtain result. In this study, the parameters of uric acid examination were chosen because uric acid is a rheumatic disease that ranks third after arthrosis and rheumatoid arthritis. It was estimated that almost 80% of the population aged 40 years or older have gout sufferers in Indonesia. Therefore, a study was conducted on the accuracy and precision of the POCT method and the uricase enzymatic colorimetric method, which aims to determine the accuracy and precision of uric acid examination POCT method and uricase enzymatic colorimetric method. The data analysis used is descriptive analysis to analyze the value of accuracy and precision. As the results of this research, 12.30% of bias value and 110.88% of recovery value were generated. While the coefficient of variation from the POCT method were 3.5% and the uricase enzymatic colorimetric method were 5.37%. Based on these results, it could be concluded that the accuracy and precision of uric acid examination using the POCT method and the uricase enzymatic colorimetric method were equally good and both method provided trusted results.

Item Type: Article
Uncontrolled Keywords: Accuracy, Precision, uric acid, POCT, Uricase Enzymatic Colorimetry.
Subjects: R Medicine > Medical Electronics
Divisions: Poltekkes Kemenkes Surabaya > Pusat Penelitan dan Pengabdian Masyarakat > Penelitian
Depositing User: Noer Amelia
Date Deposited: 13 Apr 2023 05:45
Last Modified: 13 Apr 2023 05:45
URI: http://repo.poltekkesdepkes-sby.ac.id/id/eprint/6740

Actions (login required)

View Item View Item