Nurhuda, All Adin and Indrato, Tri Bowo and Mak'ruf, Muhammad Ridha and Luthfiyah, Sari (2022) Effect of Irradiation Distance on Tube Voltage Measurement and X-Ray Device Time Using Scintillator. Journal of Electronics, Electromedical Engineering, and Medical Informatics (JEEEMI), 4 (2). pp. 103-108. ISSN 2656-8632
Text
168-Article_Text-775-1-10-20220502.pdf.pdf Download (1MB) |
Abstract
The results of the output of the X-ray device are very important to know how much the correct value, whether it is in accordance with the arrangements made by radiology personnel or there is a difference even deviatio n of the value out of the arrangement. This conformity test activity needs a testing tool that is often used by BAPETEN personnel to find out how much the output value of KV, Time, Dose, Room leak, mA and mAs from an X-ray device unit. The purpose of this study was to analyze the effect of irradiation distance on tube voltage measurements and X- ray device time using Scintillators. The study used scintillator sensors to detect radiation, Arduino as a programming source, Bluetooth HC-05 as digital communicati on between hardware and PC, PC / Delphi as display. This research design is Pre-Experimental with After Only Design research type. Where the author takes data compared to standard tools then analyzes the data. The results showed the largest error at a distance of 120cm with a 90 KV setting of 43.52%. And the smallest error is at a distance of 120cm with a 50 KV setting of 0.07%.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | X-Ray, Scintillator, Arduino |
Subjects: | R Medicine > Medical Electronics > Clinical Engineering R Medicine > Medical Electronics |
Divisions: | Poltekkes Kemenkes Surabaya > Pusat Penelitan dan Pengabdian Masyarakat > Pubikasi |
Depositing User: | Nanik Indra Putri Sari |
Date Deposited: | 11 Apr 2023 08:43 |
Last Modified: | 17 Apr 2023 07:10 |
URI: | http://repo.poltekkesdepkes-sby.ac.id/id/eprint/6652 |