Keakurasian sensor occlusion SENO257 pada infusion device analyzer 2 channel tampil TFT

Asrori’, Ach Jiddan (2023) Keakurasian sensor occlusion SENO257 pada infusion device analyzer 2 channel tampil TFT. Skripsi thesis, Politeknik Kesehatan Kemenkes Surabaya.

[thumbnail of 1. Cover.pdf] Text
1. Cover.pdf

Download (86kB)
[thumbnail of 2. Halaman Persetujuan Pembimbing.pdf] Text
2. Halaman Persetujuan Pembimbing.pdf
Restricted to Repository staff only

Download (156kB)
[thumbnail of 3. Halaman Pengesahan Penguji.pdf] Text
3. Halaman Pengesahan Penguji.pdf
Restricted to Repository staff only

Download (153kB)
[thumbnail of 4. Abstrak.pdf] Text
4. Abstrak.pdf

Download (128kB)
[thumbnail of 5. Kata Pengantar.pdf] Text
5. Kata Pengantar.pdf
Restricted to Repository staff only

Download (196kB)
[thumbnail of 6. Daftar Isi.pdf] Text
6. Daftar Isi.pdf

Download (36kB)
[thumbnail of 7. Daftar Tabel.pdf] Text
7. Daftar Tabel.pdf

Download (30kB)
[thumbnail of 8. Daftar Gambar.pdf] Text
8. Daftar Gambar.pdf

Download (31kB)
[thumbnail of 9. Bab 1.pdf] Text
9. Bab 1.pdf
Restricted to Repository staff only

Download (113kB)
[thumbnail of 10. Bab 2.pdf] Text
10. Bab 2.pdf
Restricted to Repository staff only

Download (308kB)
[thumbnail of 11. Bab 3.pdf] Text
11. Bab 3.pdf
Restricted to Repository staff only

Download (137kB)
[thumbnail of 12. Bab 4.pdf] Text
12. Bab 4.pdf
Restricted to Repository staff only

Download (270kB)
[thumbnail of 13. Bab 5.pdf] Text
13. Bab 5.pdf
Restricted to Repository staff only

Download (157kB)
[thumbnail of 14. Bab 6.pdf] Text
14. Bab 6.pdf
Restricted to Repository staff only

Download (36kB)
[thumbnail of 15. Daftar Pustaka.pdf] Text
15. Daftar Pustaka.pdf

Download (97kB)
[thumbnail of 16. Lampiran.pdf] Text
16. Lampiran.pdf
Restricted to Repository staff only

Download (47kB)
[thumbnail of 17. Artikel.pdf] Text
17. Artikel.pdf
Restricted to Repository staff only

Download (288kB)

Abstract

Both the syringe pump and the infusion pump are used to administer fluids or medications into a blood vessel in a steady, timed stream. Infusion pumps and syringe pumps often have issues with obstruction or occlusion. Because of a blockage in the infusion system, the incoming medication fluid does not continuously flow. The ECRI-recommended upper limit for occlusion is 20 PSI. Calibration must be performed at least once a year to assure accuracy. The goal of this study is to determine how well the pressure sensor works for measuring occlusion on the Infusion Device Analyzer's second channel of visual display. This investigation has dual channels, allowing for simultaneous tool calibration. This module's design incorporates a Water Pressure Sensor for occlusion measurement and a solenoid valve for pressure modeling. When pressure is applied to the sensor, data is sent to an Arduino for analysis. After the pressure readings have been taken, they are graphed and shown in numerical form on a 7- inch TFT LCD before being saved to an SD card.

Item Type: Thesis (Skripsi)
Contributors:
Contribution
Contributor
NIDN/NIDK
UNSPECIFIED
Yulianto, Endro
nidn4017077601
UNSPECIFIED
Rahmawati, Triana
nidn4023068101
Additional Information: 11SKR20230001
Uncontrolled Keywords: Calibration, Occlusion, Syringepump, Infusepump
Subjects: R Medicine > Medical Electronics
Divisions: Poltekkes Kemenkes Surabaya > Jurusan Teknologi Elektromedik > Program Studi Sarjana Terapan Teknologi Rekayasa Elektro-medis
Depositing User: Misnawar
Date Deposited: 26 Jan 2024 01:18
Last Modified: 26 Jan 2024 01:18
URI: http://repo.poltekkesdepkes-sby.ac.id/id/eprint/7751

Actions (login required)