Measuring Respiration Rate Based on Android

Shofiyah, Shofiyah and Wisana, I Dewa Gede Hari and Triwiyanto, Triwiyanto and Luthfiyah, Sari (2019) Measuring Respiration Rate Based on Android. IJEEMI (Indonesian Journal of Electronics, Electromedical Engineering and Medical Informatics), 1 (1). pp. 20-26. ISSN 2656-8624

[thumbnail of 25-Article_Text-119-2-10-20190823.pdf.pdf] Text
25-Article_Text-119-2-10-20190823.pdf.pdf

Download (2MB)

Abstract

Abstract-Respiratory rate is the total number of breath or breathing cycle, which occurs every minute. Abnormal respiratory rate is a sensitive indicator for danger patients requiring medical treatment immediately. The objective of the study is to design a respiration rate monitor via an Android mobile phone. In this study, we used flex sensors to detect the respiration rate. The flex sensors were placed in the human stomach diaphragm which detects the changes in the human stomach diaphragm during breathing. The measurement results are displayed on the liquid crystal display (LCD) 2 x 16. The data will be sent via a Bluetooth connection to the android to display the values and graphs. The comparison between the design and standard showed that the maximum error is 4.69% while the minimum error is 1.52%. The average error for all measurement is 2.83%. It can be concluded that the tool wear is eligible because it is still below the minimum threshold of 10% error.

Item Type: Article
Uncontrolled Keywords: Respiration rate, Android, Arduino microcontroller, Bluetooth
Subjects: R Medicine > Medical Electronics > Biomedical Engineering
R Medicine > Medical Electronics > Clinical Engineering
R Medicine > Medical Electronics
Divisions: Poltekkes Kemenkes Surabaya > Pusat Penelitan dan Pengabdian Masyarakat > Penelitian
Depositing User: Nanik Indra Putri Sari
Date Deposited: 24 Aug 2021 02:12
Last Modified: 12 Apr 2023 03:47
URI: http://repo.poltekkesdepkes-sby.ac.id/id/eprint/2444

Actions (login required)

View Item View Item